5080A Oscilloscope Calibration Option

5080A Oscilloscope Calibration Option

Oscilloscope Calibration Option for the 5080A Multi-Product Calibrator

The 5080A Oscilloscope Calibration Option enables the 5080A Multi-Product Calibrator to calibrate analog and digital oscilloscopes up to 200 MHz. It is a plug-in module that fits inside the 5080A calibrator. It can be ordered factory installed with a new calibrator, or added later at a Fluke service center for an additional installation and calibration charge.

The 5080A Oscilloscope Calibration Option adds functionality to the 5080A Multi-Product Calibrator:

  • Leveled sine wave generator up to 200 MHz for verifying bandwidth
  • Square wave generator for calibrating voltage gain
  • Fast edge risetime pulse generator (< 1 ns) for checking pulse response
  • Time mark generator for verifying horizontal deflection characteristics

Operation is easy

You access the 5080A Oscilloscope Calibration Option's capabilities simply by selecting "Scope" mode on the calibrator’s front panel. All outputs are conveniently located on the calibrator's output block, to make hook-ups easy and to minimize the errors contributed by cabling.

General Specifications
DC Voltage

Ranges: 0 V to ± 2.2 V (50 Ω),  0 V to ± 33 V (1 MΩ)

 

Best 1 Year Specification: ± 0.35%

 
AC Voltage (Squarewave)

Ranges: ± 1.8 mV to ± 2.2 V p-p (50 Ω),  ± 1.8 mV to ± 105 V p-p (1 MΩ)

 

Best 1 Year Specification: ± 0.35%

 
Fast Edge

Ranges: 4.5 mV to 2.75 V p-p (50 Ω)

 

Best 1 Year Specification: <1 ns rise time

 
Leveled Sinewave

Ranges: 50 kHz to 200 MHz

 

Best 1 Year Specification: ± 1.5% flatness

 
Time Markers

Ranges:5 s to 2 ns

 

Best 1 Year Specification:

 

± 5ppm