How to Select Thermocouple Calibration Equipment: App Notes Series, 2 of 4

Document(s): 
application/pdf icon English: Thermocouple App Notes Series, 2 of 4 - How to Select Thermocouple Calibration Equipment (2.51 MB)
application/pdf icon German: Thermocouple App Notes Series, 2 of 4 - Auswahl von Geräten zur Kalibrierung von Thermoelementen (2.21 MB)
application/pdf icon French: Thermocouple App Notes Series, 2 of 4 - Guide de sélection d'équipement d'étalonnage de thermocouple (2.21 MB)
application/pdf icon Russian: Thermocouple App Notes Series, 2 of 4 - Как выбрать оборудование для калибровки термопары (2.26 MB)
application/pdf icon Spanish: Thermocouple App Notes Series, 2 of 4 - Cómo elegir un equipo de calibración de termopares (2.21 MB)

Characteristics of thermocouples

Thermocouples can be constructed from almost any material with thermoelectric properties. The mixtures of materials in thermocouples are chosen for specific behavior and applications. A thermocouple produces a voltage output change when exposed to a change in temperature. In contrast, thermistors and PRTs produce a resistance change when the sensing element is exposed to a temperature change. Table 1 contrasts the key characteristics of thermocouples compared to thermistors and PRTs.

Table 1. Comparing characteristics of thermocouples, thermistors, and PRTs

Thermocouple Thermistor PRT
Temperature range: -200 °C to 1700 °C* Temperature range: 0 °C to 100 °C Temperature range: −200 °C to 1000 °C**

Accuracy is dependent on TC type and wire tolerances

(Standard or Special Limits)***

Approximate accuracy ranges are:

Type J ± 2.2 °C or ± 0.75 % of reading (Standard)

Type S ± 0.6 °C or ± 0.75 % of reading (Special)

Accuracy: ± 0.001 °C to ± 0.01 °C Accuracy: ± 0.006 °C to ± 0.04 °C
Stable Very stable Very stable
Rugged Moderately delicate Very delicate
Low priced Moderately priced Highly priced

*Represents the combined temperature range of common TC types. A single TC cannot be used over the full listed range.

**Represents the combined temperature range of common PRTs. A single PRT cannot be used over the full listed range.

***Standard Limits of Error thermocouples use standard grade wire, are more common, and are less expensive. Special Limits of Error thermocouples are made of higher grade wire which increases accuracy but are more expensive.

Thermocouple calibration equipment

Calibration is performed by measuring the voltage output of the thermocouple unit under test (UUT) while the measuring (hot) junction is exposed to a temperature source and the reference (cold) junction is maintained at a reference (usually 0 °C). The following instruments are required:

  • Reference probe
  • Readout(s) for the reference probe and thermocouple under test
  • Temperature source for the reference probe, thermocouple under test, and thermocouple reference junction

Reference probe

Depending upon the accuracy required, the reference probe will be an SPRT, a PRT, or a thermocouple of better quality and calibration than the thermocouples under test. Since this instrument is the reference for calibration, its accuracy and stability are important.

SPRTs

SPRTs are the most accurate and stable reference probes available. Generally, a glass-sheath version is used. These instruments are standardized since they are part of the ITS-90 definition, which means there are minimum requirements for the purity of the platinum wire and the type of construction used. This results in less confusion concerning the suitability of the instrument for a particular application and confidence in its performance if calibrated and used correctly. These instruments are highly stable and accurate, but they are expensive and extremely delicate. They should be reserved for high accuracy applications only.

PRTs

When accuracy requirements are less critical, PRTs can be used successfully. PRTs are available in many configurations. However, PRTs which are suitable for use as calibration standards are generally available as stainless steel or Inconel-sheathed probes. These instruments are not as accurate as SPRTs but are generally more rugged and easier to work with. Unlike SPRTs, the design of PRTs is left to the discretion of the manufacturer. Not all designs perform to the level required for use as a reference. Use care in selecting a PRT to ensure that the type selected is appropriate for use as a calibration reference over the range of interest and with the required accuracy.

Thermocouples

Reference grade thermocouples are available with uncertainties and stability approaching a PRT (or even an SPRT) at high temperatures. This grade is suitable as a calibration standard. Reference grade thermocouples are standardized in composition but not construction. Take precautions to ensure that the model selected will perform as required over the temperature range of interest.

Special considerations

In addition to accuracy requirements, there are other characteristics which must be considered. For example, the reference junction end of the thermocouple must be long enough to allow proper immersion into the reference temperature source (typically an ice bath). Also, some of the common sheath materials used in thermocouple probes should not be used at high temperatures. Ensure that the sheath material is compatible with the calibration process for which it is intended.

Readout

Since thermocouples produce a voltage output, the requirements for a thermocouple readout are different from those for an SPRT, PRT, or thermistor readout. Unless the reference probe is a thermocouple also, two readouts will be required. The thermocouple reference junction must be considered. Most thermocouple readouts have “electronic reference junctions,” often referred to as “cold junction compensation.” This is an additional circuit, which measures the temperature at the thermocouple readout connection when connecting thermocouple wire directly to the readout, and it compensates for the non-zero reference temperature. This type of compensation is very convenient, but not usually as accurate as an actual ice point bath.

Best results will be obtained with readouts designed specifically for thermocouple calibration. DMMs severely limit the flexibility with usually no cost savings or accuracy increase.

The voltage output from a thermocouple is very low, and a small voltage uncertainty equates to a large temperature uncertainty. The voltage measurements must be extremely accurate even for moderate accuracy temperature calibrations. Also, at the low voltage levels that will be measured when calibrating thermocouples, the readout floor error (noise limit or zero offset limit) becomes very significant. Ensure that the readout has a voltage range (usually to 100 mV range) and accuracy appropriate for thermocouple calibrations. Consider an example using a high accuracy 7.5 digit DMM to measure a type S thermocouple at 500 °C. The following example shows the relative contribution from the DMM sources of error.

Thermocouple construction
Figure 1.Relative contribution from the DMM sources of error.

In this example, the uncertainty resulting from the DMM floor is much larger than the error due to the DMM range accuracy. This situation is more pronounced at lower temperatures and less pronounced at higher temperatures. This illustrates the importance of readout floor error.

Temperature source

The most common temperature sources for thermocouple calibration are dry-wells and furnaces. When even higher accuracy is required, calibration baths can be utilized. For the lowest temperatures (below –100 °C) an LN2 (liquid nitrogen) comparison device is required.

Stability and uniformity should be considered in selecting the temperature source since these factors will contribute to the thermocouple calibration uncertainty:

  • Stability specifies how precisely the temperature source maintains a set-point temperature over time.
  • Uniformity specifies how consistent the temperatures are throughout the temperature source.

Bare wire thermocouples should never be immersed directly into bath fluid. A protection tube should be used. Thermocouple probes are not usually massive, but immersion depth must still be considered. Insufficient immersion depth will result in calibration errors. At elevated temperatures, precautions must be taken to avoid damage to the reference probe. Additionally, if an external reference junction is to be used, ensure that the temperature source selected is sufficiently insulated so that the external surfaces do not get hot enough to damage the ice bath. Carefully evaluate your requirements before selecting the temperature source to ensure the instrument will match your application.

Table 2. Fluke Calibration equipment recommended for thermocouple calibration

Reference Probes
Model Range Size Basic Accuracy *
SPRTs
5698-25 –200 °C to 670 °C Quartz, 485 mm x 7 mm (19.1 in x 0.28 in) < 0.006 °C/100 hours at 670 °C
Secondary Standard PRT
5626 –200 °C to 661 °C 305 or 381 mm x 6.35 mm (12 or 15 in x 0.25 in) ± 0.007 °C at 0 °C
5615-12 –200 °C to 420 °C 305 mm x 6.35 mm (12 in x 0.25 in) ± 0.013 °C at 0.010 °C
Precision Industrial PRT
5627A-9 –200 °C to 300 °C 229 mm x 4.7 mm (9 in x 0.19 in) ± 0.05 °C at 0 °C
5627A-12 –200 °C to 420 °C 305 mm x 6.35 mm (12 in x 0.25 in) ± 0.05 °C at 0 °C
Type R and S Standard Thermocouples
5649/5650-20 0 °C to 1450 °C 508 mm x 6.35 mm (20 in x 0.25 in) ± 0.7 °C at 1100 °C
5649/5650-25 0 °C to 1450 °C 635 mm x 6.35 mm (25 in x 0.25 in) ± 0.7 °C at 1100 °C
at 1100 °C *Basic accuracy includes calibration uncertainty and short-term repeatability. It does not include long-term drift.

Table 2 continued. Fluke Calibration equipment recommended for thermocouple calibration

Readouts
Model Probe Types Accuracy Features
1523 PRTs, Thermocouples, Thermistors

± 0.015 °C at 0 °C (PRTs)

± 0.24 °C at 0 °C (Type K TC)

Battery-powered, handheld reference thermometer; INFO-CON connector reads coefficients without programming; saves 25 readings on demand; graphs trends
1524 PRTs, Thermocouples, Thermistors

± 0.015 °C at 0 °C (PRTs)

± 0.24 °C at 0 °C (Type K TC)

Handheld reference thermometer same as 1523 but with inputs for two thermometers; logs up to 15,000 readings and stores 25 more on demand
1529 PRTs, Thermocouples, Thermistors

± 0.006 °C at 0 °C (PRTs)

± 0.4 °C at 600 °C (Type K TC, int CJC)

Four channels can all be measured simultaneously; battery-powered; logs up to 8,000 readings; flexible display
1586A with DAQ-STAQ Multiplexer PRTs, Thermocouples, Thermistors

± 0.005 °C at 0 °C (PRTs)

± 0.29 °C at 0 °C (Type K TC, int CJC)

40 channels with scan rate of 10 channels per second; automates sensor calibration when connected to a Fluke Calibration temperature source
Temperature Sources
Model Range Accuracy
Field Metrology Wells
9142 -25 °C to 150 °C ± 0.2 °C
9143 33 °C to 350 °C ± 0.2 °C
9144 50 °C to 660 °C

± 0.35 °C at 50 °C

± 0.35 °C at 420 °C

± 0.5 °C at 660 °C

Metrology Wells
9170 -45 °C to 140 °C ± 0.1 °C
9171 -30 °C to 155 °C ± 0.1 °C
9172 35 °C to 425 °C

± 0.1 °C at 100 °C

± 0.15 °C at 225 °C

± 0.2 °C at 425 °C

9173 50 °C to 700 °C

± 0.2 °C at 425 °C

± 0.25 °C at 660 °C

Thermocouple Furnaces
9150 (vertical) 150 °C to 1200 °C ± 5 °C
9118A (horizontal) 300 °C to 1200 °C ± 5 °C

See the other app notes in this thermocouple series:

1 of 4: Thermocouple Fundamentals

3 of 4: Calculating Uncertainties in a Thermocouple Calibration System

4 of 4: How to Calibrate a Thermocouple

Recommended Products:

5649/5650 Type R and Type S Thermocouple Standards

9118A Thermocouple Calibration Furnace

  • Home
  • Products
    • New Products
    • Electrical Calibration
      • Electrical Standards
      • Electrical Calibrators
      • Bench Multimeters
      • Electrical Calibration Software
    • RF Calibration
      • RF Reference Sources
      • RF Calibration Accessories
      • RF Calibration Software
    • Data Acquisition and Test Equipment
      • Bench Multimeters
      • Data Acquisition
      • Data Acquisition Software
    • Temperature Calibration
      • ITS-90 Temperature Standards
        • ITS-90 Fixed-point cells
        • Standard Platinum Resistance Thermometers
        • Maintenance Apparatus
        • Liquid Nitrogen Comparison Calibrator
        • Resistance Bridges
        • Standard Resistors
      • Calibration Baths
        • Compact Calibration Baths
        • Standard Calibration Baths
        • Special Application Baths
        • Bath Accessories
        • Bath Controllers
        • Bath Fluids
      • Industrial Calibrators
        • Field Metrology Wells
        • Metrology Wells
        • Handheld Calibrators
        • Field Dry-Block Calibrators
        • Micro Baths
        • Infrared Calibrators
        • Thermocouple Furnaces
        • Dual Block Dry-Well
        • Zero-point Dry-Well
      • Probes / Sensors
        • Platinum Resistance Thermometers (PRTs)
        • Thermistors
        • Thermocouples
      • Digital Thermometer Readouts
      • Multifunction Calibrators
      • Temperature Calibration Software
    • Humidity Calibration
      • Humidity Generators
      • Humidity Data Loggers and Monitors
    • Pressure Calibration
      • Piston Gauges
        • NMI Piston Gauges
        • Absolute Piston Gauges
        • High Pressure Pneumatic Piston Gauges
        • Hydraulic Piston Gauges
        • Piston Gauge Accessories
      • Automated Pressure Controller / Calibrators
        • Low Pressure Controllers / Calibrators
        • Pneumatic Pressure Controllers / Calibrators
        • High Pressure Pneumatic Controller / Calibrators
        • Hydraulic Controller / Calibrators
      • Pressure Monitors
        • Low Pressure Monitors
        • Digital Pressure Gauges
        • Reference Pressure Monitors
      • Deadweight Testers
        • Pneumatic Deadweight Testers
        • Oil Deadweight Testers
        • Water Deadweight Testers
        • High-Pressure Hydraulic Deadweight Testers
        • Deadweight Tester Accessories
      • Manual Pressure Calibration
        • Pressure Calibrators
        • Pneumatic Pressure Control
        • Hydraulic Pressure Comparators / Pumps
      • Handheld Pressure Calibrators
      • Air Data Calibration
      • Environmental Monitors
      • Pressure Calibration Accessories
      • Pressure Calibration Custom Systems
      • Pressure Calibration Software
    • Flow Calibration
      • Gas Flow Standards
      • Gas Flow Accessories
      • GFS Primary Gravimetric Flow Standard
      • Flow Calibration Software
    • Process Calibration Tools
      • Temperature Calibrators
        • Handheld Temperature Calibrators
        • Dry-Block Calibrators and Micro-Baths
        • Precision Digital Thermometers
        • Temperature Probes
        • Infrared Calibrators
        • Hygro Thermometer with Data Logging
      • Pressure Calibrators
        • Digital Pressure Calibrators
        • Handheld Pressure Calibrators
        • Deadweight Testers
        • Precision Digital Pressure Gauges
        • Calibration Hand Pumps
      • Multifunction Calibrators
      • mA Loop Calibrators
      • Process Calibration Software
    • Calibration Software
      • MET/CAL® Software
      • MET/CAL® Support
      • Asset Management Software
      • Temperature Calibration Software
      • Pressure Calibration Software
      • Flow Calibration Software
      • Mechanical / Dimensional Calibration Software
    • Service and Support
    • All Calibration Instruments
    • Handheld Test Tools
  • Purchase Info
    • Where to Buy
    • Request a Quote
    • Request a Demo
    • Request a Sales Consultation
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • News
    • Press Releases
    • Promotions
    • Industry Links
    • Metrology Salary Survey
  • Training and Events
    • Conferences and Exhibitions
    • Training Courses
    • User Group Meetings
    • "How To" Seminars
    • Web Seminars
      • Live Seminars
      • On-Demand Seminars
  • Literature and Education
    • Education Hub
    • About Calibration
    • Articles and Education
    • Product Literature
    • Product Manuals (User Guides)
    • Videos and Virtual Demos
    • Resource Centers
    • Blog
  • Service and Support
    • Service Request (RMA)
    • Service Plans
    • Technical Support
    • Knowledge Base
    • Accreditations
    • Authorized Service Centers
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • Product Manuals (User Guides)
    • Safety Data Sheets (SDS)
    • Safety, Service, and Product Notices
    • Software Downloads
    • Tools
      • PT100 Calculator
      • PT100 Table Generator
      • ITS-90 Reference Function Calculator
      • Thermocouple Table Voltage Calculator
      • Thermocouple Voltage to Temperature Calculator
  • About Us
    • Contact Us
    • Visit Us
    • Employment
    • Fluke Calibration
      • Hart Scientific Temperature Calibration
      • DH Instruments Pressure and Flow Calibration
      • Pressurements Pressure Calibration
      • Ruska Pressure Calibration
    • Fluke Companies
    • Integrity and Compliance
    • Frequently Asked Questions
    • Why buy Fluke Calibration ?
Home|Literature|Articles and Education|Temperature Calibration|Application Notes|How to Select Thermocouple Calibration Equipment: App Notes Series, 2 of 4

Products

  • Product index
  • New products
  • Where to buy
  • Request a quote

Services

  • Calibration and repair
  • About Fluke Calibration
  • Service centers
  • Software downloads

Resources

  • Catalogs
  • Manuals
  • Web seminars
  • Subscribe to e-news
  • Supplier Handbook

Company

  • Contact us
  • About Fluke Calibration
  • Why choose us
  • Other Fluke companies
©1995-2023 Fluke Corporation
             

Secondary menu

  • Disclaimer
  • Privacy Statement
  • Terms of Use
  • Terms of Sale