- Other Fluke companies:
- Fluke
- Fluke Biomedical
- Fluke Networks
- Fluke Process Instruments
- Pressure ranges to 10,000 psi (700 bar)
- 0.015% of reading accuracy standard (0.008% optional)
- Single or dual piston formats
- Built-in hand pumps standard
- Mounted spirit level with adjustable feet
- Units can be trimmed to local gravity FOC
The P3200 Models are available in single or dual piston formats for increased operating ranges. Units can be supplied in psi, bar, kgf/cm², and MPa. These robust instruments are highly accurate, quick and easy to use. Units feature a built-in priming pump for large volume applications, piston flotation indicators and a high quality screw press for fine pressure control.
Optional PressCal software is available for pressure calculations and certificate generation.
Operating principle
Deadweight testers are the high performance pressure standard for pressure measurement. Utilizing the well proven piston-gauge system consisting of a vertically mounted, precision lapped piston and cylinder assembly, accurately calibrated weight masses (Force) are loaded on the piston (Area), which rises freely within its cylinder. These weights balance the upward force created by the pressure within the system. The pressure is measured when placed on a correctly spinning and floating piston. The total pressure measured is the summation of the weights plus the piston weight carrier assembly.
There are three basic variations in the hydraulic offering; single low pressure, single high pressure and dual piston models. Pressure is generated and controlled by means of a high quality screw press located on the front of the tester. A built-in hand pump is included as standard for all hydraulic models to prime the system and accommodate large volume requirements.
The piston/cylinder assembly is the heart of each deadweight tester. They are manufactured from materials that provide stability, durability, and low thermal coefficients and distortion. Our experience and knowledge of piston/cylinder production and calibration ensure the precision and performance required for today’s demanding calibration requirements.
Standard weight masses are series 3 nonmagnetic austenitic stainless steel. Each mass is marked with the serial number of the instrument and the nominal pressure value relative to the high or low-pressure piston, when applicable. Optional fractional weights are stainless steel and/or solution heat treated aluminum.
Gravity varies significantly with geographical location and this variation has a direct effect on the force of the weights and the accuracy of the deadweight tester. Each instrument can be calibrated to local gravity at no extra cost. If unspecified, instruments will be calibrated to Standard Gravity at 980.665 cm/s2.
Pressure Ranges |
|
Distilled water operated | Ranges to 10,000 psi (700 bar) |
Accuracy |
|
Accuracy | ±0.015% of reading (±0.008% optional)
*Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. |
Materials of Construction |
|
Standard weight material | Series 3 non-magnetic austenitic stainless steel |
Weight density | 7.8 g/cm³ |
Optional fractional weights | Solution heat treated aluminum |
Weight density | 2.7 g/cm³ |
Piston material | Tungsten carbide with nickel binder Density- 15.0 g/cm³ |
Cylinder material | Hardened martensitic steel Tungsten carbide (water above 500 psi, 35 bar) |
Thermal coefficients of expansion | Oil piston/cylinder 16.5 ppm/°C Water (above 500 psi, 35 bar) 11 ppm/°C |
General |
|
Test port adaptors | 1/8, 1/4, 3/8 and 1/2 NPT or BSP |
Instrument weight | 36 lbs (16 kg) |
Instrument size | (W x D x H) 17.5 in x 12 in x 8.5 in (440 mm x 300 mm x 215 mm) |
Mass set weight (typical) | 80 lbs (36 kg) |
Reservoir volume | 150 cc |
Screw press displacement | 5.5 cc |
Pump displacement | 4.7 cc per stroke |
O-ring seal materials | Buna N as standard, Viton and EPDM available |
Weight Increments |
|
Minimum Standard Weight Increments | Up to 500 psi (35 bar): 1 psi (0.1 bar) 500 to 5000 psi (35 to 350 bar): 10 psi (1 bar) 5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar) Model P3112: 1 psi (0.1 bar) |
Optional Fractional Weights | Up to 500 psi (35 bar): 0.1 psi (0.01 bar) 500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar) 5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar) |
Operating Fluids |
|
Operating Fluids | Distilled or deionized water |
Options |
|
PressCal Software | Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments |
Fluids | Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory. |
Model Name | Description |
---|---|
P3211-BAR |
Hydraulic Deadweight Tester |
P3211-KGCM2 |
Hydraulic Deadweight Tester |
P3211-PSI |
Hydraulic Deadweight Tester |
P3211-KPA |
Hydraulic Deadweight Tester |
P3211-MPA |
Hydraulic Deadweight Tester |
P3213-BAR |
Hydraulic Deadweight Tester |
P3213-KGCM2 |
Hydraulic Deadweight Tester |
P3213-PSI |
Hydraulic Deadweight Tester |
P3213-MPA |
Hydraulic Deadweight Tester |
P3214-BAR |
Hydraulic Deadweight Tester |
P3214-KGCM2 |
Hydraulic Deadweight Tester |
P3214-PSI |
Hydraulic Deadweight Tester |
P3214-MPA |
Hydraulic Deadweight Tester |
P3223-BAR |
Hydraulic Deadweight Tester |
P3223-KGCM2 |
Hydraulic Deadweight Tester |
P3223-PSI |
Hydraulic Deadweight Tester |
P3223-MPA |
Hydraulic Deadweight Tester |
P3224-BAR |
Hydraulic Deadweight Tester |
P3224-KGCM2 |
Hydraulic Deadweight Tester |
P3224-PSI |
Hydraulic Deadweight Tester |
P3224-MPA |
Hydraulic Deadweight Tester |
Accessories common to all models:
Accessory | Description |
---|---|
P5521 and P5522 Liquid-to-Liquid Separators |
These liquid-to-liquid separators connect directly to the test port of a hydraulic deadweight tester or comparison test pump. A flexible diaphragm separates the fluids, protecting the calibrator from contamination, and allows calibration of the device in its specific working fluid. |
P5523 Liquid-to-Gas Media Separator |
Pressure range 0 to 3,000 psi (210 bar) For high pressure pneumatic calibrations this unit interfaces with hydraulic deadweight testers providing a rateless liquid-to-gas separation. Driven by a pneumatic pressure source (nitrogen bottle) and controlled by needle valves, high pressure gas is balanced against the liquid in a hydraulic deadweight tester through a series of fluid traps. * Note---since fluid mist may be transferred during operation we do not recommend this device for oxygen safe instruments using an oil operated deadweight tester.
|
P5543 Angle Adapter |
To calibrate gauges with the pressure connection on the rear (for example, panel mount gauges) in their correct operating position, an angle adaptor should be used. The angle adapter uses the standard gauge adapters and positions the gauges at 90º. The maximum working pressure of this unit is 10,000 psi (700 bar). |
P5544 Two Gauge Stand |
This adapter mounts directly to the test port of the calibrator and allows for the calibration of two instruments at the same time, or the connection of a reference test instrument. Maximum working pressure is 10,000 psi (700 bar). |
P5551 Pointer Remover / Punch |
This tool is designed to quickly remove and consistently refit the pointer of a pressure gauge. |
Product Manuals |
---|
Brochures |
---|
Application Notes |
---|
Data Sheets |
---|
Papers / Articles |
---|
Performing Precise Pressure Calibrations May Cost Less Than You Think |
Video |
---|
Video Tutorial: How to measure pressure with a deadweight tester? |
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Handheld Test Tools
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us
Sidebar Request a Quote
Request a quote