- Other Fluke companies:
- Fluke
- Fluke Biomedical
- Fluke Networks
- Fluke Process Instruments
See more Fluke brands
Standard Platinum Resistance Thermometer
The 8508A-SPRT Extended Range Metal-Sheath SPRT has been designed and manufactured by Fluke Calibration's primary standards metrologists, to complement the Fluke 8508A Reference Multimeter’s precision PRT temperature function.
Specifications | |
Temperature Range | –200 °C to 670 °C |
Nominal RTPW | 25.5Ω (± 0.5Ω) |
Current | 1 mA |
Resistance Ratio | W(302.9146 K) ≥ 1.11807 W(234.3156 K) ≤ 0.844235 |
Sensitivity | 0.1 Ω/ °C |
Drift Rate | < 0.008 °C/year (< 0.003 °C/year typical) |
Repeatability | < 1 mK |
Self-heating at TPW | < 0.001 °C under 1 mA current |
Reproducibility | ± 0.001 °C or better |
RTPW Drift After Thermal Cycling | < 0.001 °C |
Diameter of Pt Sensor Wire | 0.07 mm (0.003 in) |
Lead Wires | Four sensor wires plus grounding wire |
Protective Sheath | Inconel |
Diameter: | 5.56 mm ± 0.13 mm (0.219 in ± 0.005 in) |
Length: | 482 mm (19 in) |
Insulation Resistance | > 100 MΩ at 661 °C > 1000 MΩ at 20 °C |
Calibration |
Accredited calibrations from Fluke Calibration included |
Request a quote