8508A-PRT
Platinum Resistance Thermometer
Highlights
- Range to 661°C
- Meets all ITS-90 requirements for resistance ratios
- Rtp drift < 20 mK after 500 hours at 661°C
- Calibrated accuracy of ± 0.006 °C at 0 °C
The 8508A-PRT Secondary Standard PRT Temperature Sensor has been designed and manufactured by Fluke Calibration's primary standards metrologists, to complement the Fluke 8508A Reference Multimeter’s precision PRT temperature function.
Specifications |
|
Temperature Range |
–200 °C to 661 °C |
Handle Temp. |
0 °C to 80 °C |
RTPW |
100 Ω (± 1 Ω) |
Resistance Ratio W(Ga) |
W(302.9146K) ≥ 1.11807 |
Calibrated Accuracy† (k=2) |
± 0.006 °C at –200 °C |
Stability |
± 0.003 °C |
Long-Term Drift (k=2) |
< 0.006 °C/100 hours at 661 °C |
Immersion |
At least 12.7 cm (5 in) recommended |
Sheath |
Inconel™ 600 |
Lead Wires |
4-wire Super-Flex PVC, 22 AGW |
Size |
6.35 mm dia. 381 mm (0.25 x 15 in) |
Calibration |
Accredited calibrations from Fluke Calibration included |
†Includes calibration and 100 hr drift |
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Handheld Test Tools
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us
Sidebar Request a Quote
Request a quote