Key features

  • Highly accurate, full-featured , easy-to-use single function RTD temperature calibrator
  • Measures and simulates 13 different RTD types and resistance
  • Measures 4 to 20 mA signals while simultaneously sourcing a temperature signal
  • Comes with a magnetic hanging tool designed in for convenient hands-free operation
  • Features configurable 0% and 100% source settings for quick 25% linearity checks
In stock
$1,970.99
FREE Shipping
30-day Returns

Product overview: Fluke 712B RTD Calibrators

Fluke 712B RTD Calibrator provides highly accurate, easy-to-use, single-function RTD temperature calibration

Fluke 712B RTD Calibrator is a handheld, battery-operated instrument that measures and sources a variety of RTD types and resistances. It also includes an isolated channel to measure 4-20 mA whiles sourcing a temperature signal. It offers configurable 0% and 100% source settings for quick 25% linearity checks. It also provides linear ramp and 25% step auto ramp functionality based on 0% and 100% settings.

Add to that dual inputs, an easy to view backlit display, plus the ability to memorize settings at power down for easy restart make the 712B the ideal test tool for temperature calibration professionals who want a highly accurate, easy-to-use, single function RTD temperature calibrator.

Other useful features:

  • Dual inputs and backlit display for easy interpretation of measurements
  • Remembers power down settings to easily restart tests when powered up
  • Provides linear ramp and 25% step auto ramp based on 0% and 100% settings
  • Provides one-year and two-year specifications and traceable certification of calibration

Specifications: Fluke 712B RTD Calibrators

General Specifications
Maximum voltage applied between any terminal and earth ground or between any two terminals30 V
Operating temperature-10°C to 50°C
Storage temperature-30°C to 60°C
Operating altitude2,000 m
Storage altitude12,000 m
Relative humidity (% RH operating without condensation)Non condensing
90% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
(Without condensation)
Vibration requirementsMIL-T-28800E, Class 2
Drop text requirements1 m
IP RatingIEC 60529: IP52
Electromagnetic environmentIEC 61326-1, Portable
SafetyIEC 61010-1, Max 30 V to earth, Pollution Degree 2
Power supply4 AA NEDA 1.5A IEC LR6 batteries
Size (H x W x L)52.5 x 84 x 188.5 mm
Weight515 g
DC mA Measurement
Resolution0-24 mA
Range0.001 mA
Accuracy (% of reading + counts)0.010% + 2 μA
Temperature coefficient± (0.002% of reading + 0.002% of range) /°C ( < 18°C or >28°C)
Ohms Measurement
Ohms rangeAccuracy (% of reading +counts)
0.00 Ω to 400.00 Ω0.015% + 0.05 Ω
400.0 Ω to 4000.0 Ω0.015% + 0.5 Ω
Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω), 0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications.
Temperature coefficient±(0.002% of reading + 0.002% of range) / °C (< 18°C or > 28°C)
Ohms Source
Ohms range1.0 Ω to 400.0 Ω
1.00 Ω to 400.00 Ω
400.0 Ω to 1500.0 Ω
1500.0 Ω to 4000.0 Ω
Excitation current from measurement device0.1 mA to 0.5 mA
0.5 mA to 3 mA
0.05 mA to 0.8 mA
0.05 mA to 0.4 mA
Accuracy (% of reading + counts)0.015% + 0.1 Ω
0.015% + 0.05 Ω
0.015% + 0.5 Ω
0.015% + 0.5 Ω
Resolution0.00 Ω to 400.00 Ω0.01 Ω
400.0 Ω to 4000.0 Ω0.1 Ω
Temperature coefficient± (0.002% of reading + 0.002% of range) /°C ( <18°C or >28°C)
Supports pulsed transmitters and PLCs with pulse times as short as 5 ms
RTD Input and Output
RTD Type (α)Range (°C)Measure (°C)Source (°C)
1-year2-yearSource current1-year2-year
10 Ω Pt (385)-200 to 100°C1.5°C3°C1 mA1.5°C3°C
100 to 800°C1.8°C3.6°C1 mA1.8°C3.6°C
50 Ω Pt (385)-200 to 100°C0.4°C0.7°C1 mA0.4°C0.7°C
100 to 800°C0.5°C0.8°C1 mA0.5°C0.8°C
100 Ω Pt (385)-200 to 100°C0.2°C0.4°C1 mA0.2°C0.4°C
100 to 800°C0.015%
+0.18°C
0.03%
+0.36°C
1 mA0.015%
+0.18°C
0.03%
+0.36°C
200 Ω Pt (385)-200 to 100°C0.2°C0.4°C500 μA0.2°C0.4°C
100 to 630°C0.015%
+0.18°C
0.03%
+0.36°C
500 μA0.015%
+0.18°C
0.03%
+0.36°C
500 Ω Pt (385)-200 to 100°C0.3°C0.6°C250 μA0.3°C0.6°C
100 to 630°C0.015%
+0.28°C
0.03%
+0.56°C
250 μA0.015%
+0.28°C
0.03%
+0.56°C
1000 Ω Pt (385)-200 to 100°C0.2°C0.4°C250 μA0.2°C0.4°C
100 to 630°C0.015%
+0.18°C
0.03%
+0.36°C
250 μA0.015%
+0.18°C
0.03%
+0.36°C
100 Ω Pt (3916)-200 to 100°C0.2°C0.4°C1 mA0.2°C0.4°C
100 to 630°C0.015%
+0.18°C
0.03%
+0.36°C
1 mA0.015%
+0.18°C
0.03%
+0.36°C
100 Ω Pt (3926)-200 to 100°C0.2°C0.4°C1 mA0.2°C0.4°C
100 to 630°C0.015%
+0.18°C
0.03%
+0.36°C
1 mA0.015%
+0.18°C
0.03%
+0.36°C
10 Ω Cu (427)-100 to 260°C1.5°C3°C1 mA1.5°C3°C
120 Ω Ni (672)-80 to 260°C0.15°C0.3°C1 mA0.15°C0.3°C
50 Ω Cu (427)-180 to 200°C0.4°C0.7°C1 mA0.4°C0.7°C
100 Ω Cu (427)-180 to 200°C0.2°C0.4°C1 mA0.2°C0.4°C
YSI40015 to 50°C0.2°C0.4°C250 μA0.2°C0.4°C
1. Sensor inaccuracies not included.
2. Resolution: 0.1°C.
3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0°C (Pt10 and Cu10), 0.6°C (Pt50 and Cu50), 0.4°C (Other RTD types) to the specifications.
4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range).
5. Temperature Coefficient: ±0.05°C /°C for measure, ±0.05°C /°C (< 18°C or > 28°C) for source.
6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms.

Models: Fluke 712B RTD Calibrators

Fluke 712B
Fluke 712B RTD Calibrator
Part#: 4387857
UPC#: 0 95969 68640 2

Includes:

  • Magnetic hanging tool
  • Batteries
  • Manual
  • Traceable calibration certificate
  • Test Leads
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 24.84ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 25.44ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.
Welcome to Fluke's Chat GPT Assistant. As with other generative AI systems, this chatbot may occasionally generate incorrect information. For questions regarding the safe use of Fluke tools, please refer to your owners manual. You must comply with your employer’s safety standards and obtain necessary training before using Fluke tools or taking electrical measurements. To speak with a person, visit our Contact Us page.
How can I help you today?