7060/7080 Really Cold Temperature Calibration Baths

7060/7080 Really Cold Temperature Calibration Baths
7060/7080 Really Cold Temperature Calibration Baths

Chill to –40, –60, or –80 °C without external coolants

  • Self-contained refrigeration—no LN2 or chiller required
  • Temperatures as low as –80 °C in real metrology baths
  • Best stability and uniformity available at –80 °C and below
  • Large working areas for increased throughput

Find a refrigerated calibration bath for extremely low temperatures (i.e. –80 °C). Fluke Calibration metrology baths are well known world-wide for excellent temperature control that maintains excellent temperature, stability (±0.0015 °C) and uniformity (±0.003 °C). Models 7060 and 7080 can be used with Fluke Calibration bath fluids to reach extremely cold temperatures.

Do you need a bath that chills below –40 °C to temperatures as low as –60 °C or even –80 °C? Would you like a bath that reaches those temperatures without using any external coolants? Fluke Calibration the baths that meet these temperature requirements and give you the best stability in the industry.

These baths are completely self-contained. They require no auxiliary cooling fluids or devices to achieve their set-point temperatures. Using Fluke Calibration's unique "heat-port" design, stability at –80 °C is ±0.0025 °C. No other company makes a bath that can match a Fluke Calibration bath’s performance, and Fluke Calibration baths are backed by our guarantee that if they don’t perform exactly the way we say they will, we’ll take them back. No arguments. No ifs, ands, or buts.

If you want to completely automate the entire calibration process, see the description of Fluke Calibration's MET/TEMP II software package.

Forget commodity-like utility baths! They’re not designed for high performance calibration needs. And be careful of companies that advertise performance specifications they don’t meet. It’s easy to write down numbers; it’s more difficult to meet them with an instrument.

Remember, if our baths don’t perform the way we say they will, just send them back. Our equipment won’t disappoint you.

Specifications

7060

7080

Range −60 °C to 110 °C −80 °C to 110 °C
Stability ±0.0025 °C at −60 °C (methanol)
±0.002 °C at 0 °C (methanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
±0.0025 °C at −80 °C (methanol)
±0.0015 °C at 0 °C (methanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
Uniformity ±0.005 °C at −60 °C (methanol)
±0.005 °C at 0 °C (methanol)
±0.003 °C at 25 °C (water)
±0.005 °C at 100 °C (oil 5012)
±0.007 °C at −80 °C (methanol)
±0.005 °C at 0 °C (methanol)
±0.003 °C at 25 °C (water)
±0.005 °C at 100 °C (oil 5012)
Temperature Setting Digital display with push-button data entry
Set-Point Resolution 0.01 °C; high-resolution mode, 0.00007 °C
Display Resolution 0.01 °C
Digital Setting Accuracy ±1 °C
 
Digital Setting Repeatability ±0.01 °C
 
Heaters 500 and 1000 Watts
Access Opening 127 x 254 mm (5 x 10 in)
Depth 305 mm (12 in)
Wetted Parts 304 stainless steel
Power 230 VAC (±10 %), 50 or 60 Hz, 13 A, single phase, specify frequency
Volume 27 liters (7.2 gallons)
Weight 159 kg (350 lb.)
Size (HxWxD)
 
1168 x 775 x 483 mm (46 x 30.5 x 19 in)
 
Automation Package
 
 
Interface-it software and an RS-232 computer interface is included for setting the bath temperature via an external computer. For IEEE-488, add 2001-IEEE.
Model Name Description
7060

Standard Bath, –60 °C to 110 °C

AccessoryDescription
2016-7060

Fluid Level Adapter, 7060

7080

Standard Bath, –80 °C to 110 °C

AccessoryDescription
2016-7080

Fluid Level Adapter, 7080

Accessories common to all models:

Accessory Description
2010

Access Cover, 127 x 254 mm (5 x 10 in), Lexan

2007

Access Cover, 127 x 254 mm (5 x 10 in), Stainless Steel

2001-IEEE

IEEE-488 Interface

2069

8X Magnifier Scope, with mounts

SC1-SCBA

Silver CarePlan 1 YR, Std Cascade Baths

Mainframes included: 7060, 7061, 7080, 7081

SC2-SCBA

Silver CarePlan 2 YR, Std Cascade Baths

Mainframes included: 7060, 7061, 7080, 7081

SC4-SCBA

Silver CarePlan 4 YR, Std Cascade Baths

Mainframes included: 7060, 7061, 7080, 7081