- Pressure ranges from 0-1000 psig, 0.2-1000 psia
- New autofloat option
- Accuracy to 0.0010% (10 ppm) of reading for pressures to 100 psi (7 bar)
- Accuracy to 0.0027% (27 ppm) of reading for pressures >100 psi to 1000 psi (70 bar)
- Precision better than 3 ppm
- Stability better than 3 ppm per year
- Resolution 0.0001% (1 ppm) or 1 mg
- Gauge, absolute, vacuum and low pressure modes
General |
|
| Pressure range | Gauge mode: 0.2–1000 psig (14 mbar to 70 bar) Absolute mode: 0.2–1000 psia (14 mbara to 70 bara) Vacuum mode: -14.5 to 0 psig (-1000 to 0 mbar) (Vacuum pressure achieved depends on local barometric pressure) |
| Electrical power | 115/230 VAC, 50/60 Hz, 15 W |
| Temperature | Operating temperature 15 to 28°C (60 to 82°F) Storage temperature -20 to 70°C (-4 to 158°F) |
| Humidity | Operating humidity 20–75% relative humidity, noncondensing; storage humidity 0–90% |
| Pressure medium | High purity nitrogen or dry, clean air with less than 0.5 ppm hydrocarbon and less than 5 ppm H2O content, dew point less than or equal to -50 °C, and less than 50 micron particulate size. Although lower quality gas can be used, the frequency of piston/cylinder cleaning will increase. |
Performance |
|
| Precision (type A uncertainty) | Better than 3 ppm Values are reported at the 95% confidence level (2σ) |
| Long-term stability | Better than 1.5 ppm per two years Values are reported at the 95% confidence level (2σ) |
| Resolutionc | 1 ppm or 1 mg |
Piston/Cylinder Assemblies |
|
| Low Range | Nominal area: 0.52 in2 (3.4 cm2) Pressure range: 0.2–25 psi (14 mbar to 1.7 bar) Minimum autofloat pressure: 1 psi (70 mbar) Total uncertainty: 0.001% RDG or 0.00001 psi (threshold pressure 1 psi)c Materials: Piston is 440C stainless steel, cylinder is cemented tungsten carbide Thermal coefficient: 1.5E-05/°C |
| Lower Mid Range | Nominal area: 0.13 in2 (0.84 cm2) Pressure range: 1.7–100 psi (117 mbar to 7 bar) Minimum autofloat pressure: 5 psi (350 mbar) Total uncertainty: 0.001% RDG or 0.00004 psi (threshold pressure 4 psi)c Materials: Piston and cylinder are cemented tungsten carbide Thermal coefficient: 9.1E-06/°C |
| Upper Mid Range | Nominal area: 0.026 in2 (0.168 cm2) Pressure range: 2–500 psi (0.14 to 35 bar) Minimum autofloat pressure: 20 psi (1.4 bar) Total uncertainty: 0.0026% RDG or 0.0004 psi (threshold pressure 16 psi)c Materials: Piston and cylinder are cemented tungsten carbide Thermal coefficient: 9.1E-06/°C |
| High Range | Nominal area: 0.013 in2 (0.084 cm2) Pressure range: 2–1000 psi (0.14 to 70 bar) Minimum autofloat pressure: 35 psi (2.4 bar) Total uncertainty: 0.0026% RDG or 0.0004 psi (threshold pressure 16 psi)c Materials: Piston and cylinder are cemented tungsten carbide Thermal coefficient: 9.1E-06/°C |
Mass Set |
|
| Total mass | 13.2 lb (6 kg) |
| Maximum platter mass: | 2.2 lb (1 kg) |
Autofloat System |
|
| Autofloat controller | Positive shut-off controller automatically generates pressure and maintains piston float position at desired pressure.
Includes WinPrompt software. |
| Autofloat range | Low range piston/cylinder: 1–25 psi (70 mbar to 1.7 bar) Lower mid range piston/cylinder: 5–100 psi (350 mbar to 7 bar) Upper mid range piston/cylinder: 20–500 psi (1.4 to 35 bar) High range piston/cylinder: 35–1000 psi (2.4–70 bar) |
| Float positiond | Inductive sensor Float position resolution: 0.001 inch (0.002 cm) Sink rate resolution: 0.001 inch/minute (0.001 cm/min) |
| Piston temperatured | 4-wire 100Ω PRT Accuracy: ±0.1 °C Resolution: 0.01 °C |
| Laboratory Environment Monitor (LEM) | Sensor types Temperature: Thin film platinum 1000Ω RTD Humidity: Capacitive IC humidity sensor Barometric pressure: Piezoresistive, monolithic silicon pressure transducer Accuracy |
| Vacuum module | Thermopile sensor Accuracy: 10% of reading or 10 mTorrc, whichever is greater Resolution: 1 mTorr |
| Barometric reference sensor | Accuracy: Better than ±0.002 psi (0.14 mbar) per year Accuracy in low gauge mode: ±0.0004 psi (0.03 mbar) Resolution: 0.00015 psi (0.01 mbar) |
Computer Interface |
|
| Requirements | Pentium level processor, RS-232C interface, monitor, mouse or other pointing device, keyboard; program requires 2MB available hard disk space; Windows 95 or higher |
| Model Name | Description |
|---|---|
| 2465A-754 |
Gas Piston Gauge Autofloat Instrument Base with bell jar for gauge and absolute mode, with spare parts kit. 115/230 VAC selectable |
| 2465A-799 |
Mass Set with Trim Masses |
| 2465-725 |
0.2 - 25 psi (14 mbar - 1.7 bar) Piston Cylinder |
| 2465-727 |
1.7 - 100 psi (117 mbar - 7 bar) Piston Cylinder |
| 2465-729 |
2 - 1000 psi (140 mbar - 70 bar) Piston Cylinder |
| 2465-730 |
2 - 500 psi (140 mbar - 35 bar) Piston Cylinder |
Accessories common to all models:
| Accessory | Description |
|---|---|
| 2456-LEM |
Laboratory Environment Monitor (LEM) |
| 2456-VAC |
Vacuum Sensor Module |
| 3990-801 |
Manual Pressure Controller to 7 MPa (1000 psi) |
| 2465A-23 |
Lines and Fittings Kit |
| 99878-800 |
Vacuum Pump with auto-vent, 85 liter/minute, 50/60 Hz *Note: Autofloat system require vacuum pumps equipped with an auto-vent capability. |
| 2456-800 |
Piston Gauge Monitor, Single channel, with RS-232 interface and WinPrompt software. 115/230 VAC, 50/60 Hz |
| 2456-801 |
Piston Gauge Monitor, Dual channel, with RS-232 interface (2) and WinPrompt software. 115/230 VAC, 50/60 Hz |
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Temperature Calibration
- Pressure Calibration
- Flow Calibration
- Calibration Software
- Data Acquisition and Test Equipment
- Process Calibration Tools
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us